• Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
  • Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
  • Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
  • Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
  • Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
  • Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump

Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump

Product Type: NIBP Module with Pump
Applicable Machine: Efficia Dfm100
Condition: in Good Working
Warranty: 90 Days
Transport Package: Standard Export Packing
Customization:
Gold Member Since 2022

Suppliers with verified business licenses

Trading Company
  • Overview
  • Brief Description
  • Specifications
  • NBP Accuracy Test
  • Detailed Photos
Overview

Basic Info.

Model NO.
NBP1530898

Product Description

DFM100 NIBP Module 
Efficia DFM100 NIBP Module NBP1530898 with Pump
Brief Description

DFM100 NIBP Module with Pump is one of the main components of non-invasive blood pressure monitoring equipment for patients, the model number is NBP1530898. This module is a board with an air pump, which is the most important part to support the NBP function of the defibrillator. The use of non-invasive blood pressure measurement technology is more practical than invasive blood pressure in clinical practice, and it has little impact on the patient's health and will not cause body damage. So the DFM100 NIBP Module with Pump is a medical parts with complete performance and meeting the requirements of use.

 

Specifications
Brand  
Product Type NIBP Module with Pump
Model NBP1530898
Applicable Machine Efficia DFM100
Condition In Good Working
Warranty 90 Days
Delivery Date 3-5 Working Days


 

NBP Accuracy Test

To perform an NBP accuracy test, follow the steps below:
1 Enter service mode
2 From the Service Mode main menu, select NBP.
3 Display the NBP Services screen. You may hear a soft high-pitched tone, which is normal NBP pump operation.
4 Connect the NBP test kit to the DFM100
5 Squeeze the manometer pump and apply a pressure of 150mmHg.
6 Observe the pressure displayed on the screen and record this result as X1. The value should be 150mmHg ± 3mmHg.
7 Squeeze the manometer pump and apply a pressure of 250mmHg.
8 Observe the pressure displayed on the screen and record this result as X2. The value should be 250mmHg ± 3mmHg.
9 Press Stop Test to stop the test.
10 If the difference between the pressure gauge and the displayed value is greater than 3 mmHg, replace the NBP module.
11 If the results are as described, the device passed the accuracy test. Press the Main Service soft key to return to the service mode main menu.

 

Detailed Photos
Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump
Efficia Dfm100 NIBP Module Board Nbp1530898 with Pump

 

 


 

Thank you for your time, if you are interested in Efficia DFM100 NIBP Module NBP1530898 with Pumpor have any suggestions for us, please feel free to contact us. We will reply you within 12 hours, wish you a happy life.

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now