Patient Monitor Sidestream CO2 module is used for T-series and N-series patient monitors. It can be used for routine monitoring of patients to evaluate the smoothness of the patient's entire airway and breathing circuit. The oxygen supply status and gas exchange status of the patient's respiratory system can be observed. It can also understand whether the cardiopulmonary blood circulation is normal, and effectively solve some complications of mechanical ventilation. It can also use some other functions for the medical environment, which is a very powerful module.
Patient Monitor Co2 Module is based on non -color scattered infrared spectrum analysis technology (NDIR). The principle of absorption characteristics of infrared rays is used to measure the CO2 concentration. Since the CO2 gas can absorb the infrared light of a specific wavelength, CO2 will be transmitted to a sample room, and then the infrared filter will select an infrared light with a specific frequency band through the gas. In a given volume, the higher the gas concentration, the more infrared light absorbed, and the less transmission of infrared light through the gas. The module can calculate the concentration of the gas by measuring the volume of infrared light.
Environmental Specifications
Item | Operating conditions | Storage conditions |
Barometric (mmHg) | 430 to 790 | 430 to 790 |
Relative humidity (noncondensing) | 15% to 95% | 10% to 95% |
Temperature (ºC) | 5 to 40 | -20 to 60 |
Physical Specifications
Weight | <0.48 kg |
Size | 136.5×80.5×102 mm |
Please kindly contact me, if you are interest in T series N Series Patient Monitor Sidestream CO2 module. Thank you.