• Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
  • Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
  • Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
  • Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
  • Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
  • Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002

Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002

Customized: Non-Customized
Brand: Ge
Item Name: Display Assembly
Color: White
Compatible Models: Dash3000 Dash4000 Dash5000
Transport Package: Standard Export Packing
Customization:
Gold Member Since 2022

Suppliers with verified business licenses

Trading Company
  • Overview
  • Brief Description
  • Remove&Replace
  • Q&A
Overview

Basic Info.

Model NO.
004-0004-004 2016792-002

Product Description

GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly
GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002 REV A

 
Brief Description

GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002 REV As is the display assembly of GE Dash3000 Dash4000 Dash5000 patient monitor, with internal circuit board and external frame and display. There are also button components and knobs on the frame, which are the standard equipment that a standard GE Dash3000 Dash4000 Dash5000 patient monitor should have. And the internal components and circuit boards are covered by the outer layer of metal, which is very safe and effective.
 

 
Remove&Replace
Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
1. Remove the handle assembly from the patient monitor (this step is not required for Dash 5000 patient monitors)

2. Place the patient monitor face down on a non-abrasive, static-free surface. Make sure the Trim Knob control hangs off the edge of the surface to avoid damage.
3. Use a screwdriver to remove the four screws (six screws for Dash 5000 patient monitors) securing the display assembly to the main unit.

4. Partially separate the display assembly from the main unit and look inside the unit to determine how the display flex cable is folded. Depending on the model, the Dash patient monitor may have a display flex cable that folds up from the bottom or right side of the display. (The location of the fold determines how to position the display assembly for removal from the main unit.)



 
Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
Ge Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002
5. Remove the two screws that connect the flexible circuit to the CPU/ battery housing assembly of the main unit. Pull the flexible connector stress relief device and remove the flexible connector.

6. Lift the buckle and tongue and remove the DAS connector. Place the display component face down on a non-electrostatic, non-abrasive surface.

7. Replace faulty display components.

8. Replace the Display Flex component.

 

 

CAUTION :
CABLE DAMAGE - Do not kink, pinch, stretch, twist or tightly fold the display flex cable. If the flex cable is not handled with care, it can damage the cable to the point of affecting the display integration and even the use of the entire device.

 

 

Q&A
Q: Do you guarantee the quality of your products?
A: Please rest assured that our products are professionally inspected and tested by senior engineers.
Q: May I have a look at the actual picture taken first?
A: Sure. We'll get it ready for you right away.
Q: Could you explain it to me in detail?
A: Yes, please feel free to contact us.






Please feel free to contact me, if you are interest in GE Dash3000 Dash4000 Dash5000 Patient Monitor Display Assembly 004-0004-004 2016792-002 REV A

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now